Excellent Performance

Automatic in-machine hemolysis for glycated hemoglobin
More accurate and efficient

Post-spectrometry system
Low noise, fast response, and elimination of interference from other light paths

Minimum reaction volume as low as 80μL
Improve test utilization efficiency and reduce detection costs

Multi-track collaborative system
Three-track collaboration of routine, emergency, and recycling. Independent emergency sample loading, with results available in as fast as 10 minutes

Automatic avoidance of cross-contamination
10-stage cleaning technology for reaction cups
The carry-over contamination rate is far better than the industry standard

Modular combination and pipeline expansion
Up to four-module online connection, and can also be expanded into different pipelines to easily handle the peak detection pressure in large-scale laboratories
Product Features

System Advantages
- Automatic working mode and optimized test configuration
- High-precision positioning system to ensure accurate and reliable detection
- Built-in quality control function to automatically monitor the system status

Detection Performance
- Single - module biochemisty testing speed: 2000 tests per hour
- Electrolyte analysis unit: 900 tests per hour
- Up to 60 types of tests (including ISE) can be loaded simultaneously

Information Management
- Operation interface: multilingual operation interface with user - friendly design
- Monitoring function: Monitoring of the reaction process and abnormal alerts
- Connectivity: Can be connected to the LIS system for easy data management

Reaction System
- Reaction volume: 80-360μL
- Reaction cups: Quartz cuvettes
- Reaction disk: 418 reaction positions
- Temperature control method: Solid direct heating

Sample Handling
- Sample loading method: Orbital cyclic sample loading
- STAT samples: Priority sample loading and detection
- Sample dispensing volume: 2-30μL (0.1μL step)

Optical System
- Light source: Tungsten-halogen lamp
- Spectrometry system: Holographic concave flat-field grating, post-spectrometry technology
- Measurement wavelengths: 340-850nm, a total of 14 wavelengths
For more information : wtexport@ystwt.com